Sample
Mouse serum, plasma, cell culture supernatant, or other biological samples
Intended Use
This kit is intended for the in vitro quantitative determination of GDF15 concentrations in mouse serum, plasma, cell culture supernatant and other biological samples. For Research Use Only. Not for use in diagnostic procedures.
Storage
2 - 8 °C (for sealed box), do not freeze. See kit label for expiry date.
Performance Characteristics
Perform the stability test for the sealed kit at 37 °C and 2 - 8 °C and get relevant data.
| ELISA Kit (n=5) | 37 °C for 1 month | 2 - 8 °C for 6 months | 2 - 8 °C for 12 months |
| Average (%) | 80 | 95 - 100 | 85 - 98 |
Precision
Intra-assay Precision: samples with low, medium and high concentration are tested 20 times on same plate.
Inter-assay Precision: samples with low, medium and high concentration are tested 20 times on three different plates.
| Item | Intra-assay Precision | Inter-assay Precision |
| Sample | 1 | 2 | 3 | 1 | 2 | 3 |
| n | 20 | 20 | 20 | 20 | 20 | 20 |
| Mean (pg/mL) | 32.09 | 122.2 | 481 | 32.32 | 125.2 | 492 |
| Standard deviation | 1.53 | 6.2 | 24.63 | 1.54 | 6.32 | 25.19 |
| CV (%) | 4.78 | 5.07 | 5.12 | 4.76 | 5.05 | 5.12 |
Detection Range
15.625 - 1000 pg/mL
Detection Limit
9.375 pg/mL
General Description
This kit is used for the in vitro quantitative determination of GDF15 concentrations in mouse serum, plasma, cell culture supernatant and other biological samples.
Product Features
Application: In vitro quantitative determination of GDF15 concentrations in serum, plasma, cell culture supernatant and other biological samples.
Reactivity: Mouse
Detection Method: Sandwich
Detection Range: 15.625 - 1000 pg/mL
Sensitivity: 9.375 pg/mL
Detection Duration: 4 hours (excluding balancing and sample preparation)
Samples needed for single well (Max): Serum: 50 μL, Plasma: 50 μL, Cell Culture Supernatant: 100 μL, cell or tissue lysate: 100 μL, Other liquid samples: 50 μL
Specificity: Specifically recognize GDF15, no obvious cross reaction with other analogues
Storage: 2 - 8 °C (for sealed box), please do not freeze! See kit label for expiry date
Reconstitution And Storage
The sealed kit can be stored at 2 - 8 °C. Do not freeze. For the storage condition of each component after opening, please refer to the kit components table. Store the reconstituted zero tube standard at 2 - 8 °C and use it within 12 h. Other diluted standard working solutions should be used within 2 h.