The analytical sensitivity of the Rotavirus ELISA was determined with the serial dilution of a stool sample quantified by immunoelectron microscopy (IEM). The measurements were taken in triplicate, based on a virus titer of 10.6 x 10
7 particles/ml. The detection limit was defined as 6.63 × 10
7 virus particles/ml of the stool sample. Results of the titration series are shown in Table.
Note that the positive OD value in ELISA is caused not only by intact virus particles, but also by fragments of the virus, which are not counted in the IEM.
Table: Determination of the analytical sensitivity of Rotavirus ELISA