Typical Data
The results shown below are for illustration only and should not be used to interpret results from another assay.

Sensitivity
The sensitivity, or limit of detection, of this assay is 0.59 ng/mL. It was determined by interpolation at 2 standard deviations above the mean signal at background, using data from 5 standard curves.
Specificity
This assay detects human, mouse, rat, and bovine αB-Crystallin. Cross reactivity with γ-Crystallin and αA-Crystallin is 0.06% and 0.4%, respectively. There is no cross reactivity observed with ?βL-Crystallin, Hsp10, Hsp25 and Hsp27.
Dilutional Linearity
To determine possible interference from the sample matrix, the indicated sample types were serially diluted into assay buffer. The concentrations of αB-Crystallin were measured in the assay, and the results were analyzed to determine the range over which a linear response was obtained. These data may be used as a guideline to determine minimal recommended dilution (MRD) for similar samples.

CL: Cell Lysate
Parallelism
Dose-response curves from cell lysates diluted into assay buffer (using the MRD) were compared to the recombinant αB-Crystallin standard curve. Parallelism indicates that the antibody-binding characteristics of the native and standard proteins are similar, allowing accurate determination of the analyte.