##Sample Data##
The standard curve presented here is an example of the data typically produced with this kit; however, your results will not be identical to these. You must run a new standard curve. Do not use the data below to determine the values of your samples. Your results could differ substantially.
Raw Data Average Corrected
Total Activity 0.499 0.533 0.516
NSB 0.007 0.007 0.007
B0 1.224 1.254
1.258 1.179 1.237 1.230

Typical results

Typical standard curve
Assay Range = 15.6-2,000 pg/ml
Sensitivity (defined as 80% B/B0) = 30 pg/ml Mid-point (defined as 50% B/B0) = 100-180 pg/ml
The sensitivity and mid-point were derived from the standard curve shown above. The standard was diluted with ELISA Buffer.
##Precision##:
The intra-assay CVs have been determined at multiple points on the standard curve.

Intra-assay variation
*%CV represents the variation in concentration (not absorbance) as determined using a reference standard curve.

Sample Inter-assay variation for plasma, serum, and urine
Plasma, serum and urine samples from different individuals containing high, medium, or low level of aldosterone were measured eight times on four different days using a different set of reagents each time. The calculated %CV is reported as inter-assay variance.
##Cross Reactivity##:

Cross Reactivity of the Aldosterone ELISA