1. Single-point quantification with the 4PL methodOptimized assignment of extinction signals to quantitative values is guaranteed by using non-linear functions, which adjust a sigmoide curve without any further transformation to OD-values.
Determination of antibody concentrations with the ELISA is carried out by the logistic-log-model (4 PL; 4 parameter) which is ideal for exact curve-fitting. It is based on the formula:

The parameters A, B, C, and D are representative for the exact shape of the curve:
lower asymptote → parameter A slope of the curve → parameter B turning point → parameter C upper asymptote → parameter D For each lot the standard curve is evaluated by Creative Diagnostics in several repeated test runs under optimal conditions. Time consuming and cost intensive construction of the standard curve by the user is not necessary.
For evaluation of antibody concentrations a lot specific standard curve as well as a lot specific evaluation table is included with each test kit. Appropriate evaluation software is available on request.
To compensate for normal test variations and also for test run control a standard serum is used in each individual test run. For this control serum a "reference value" with a validity range is determined by the quality control of the producer. Within this range a correct quantification of antibody concentration is ensured. Since the standard serum is not necessarily a positive control, the value of the standard serum may be borderline or negative in some ELISA tests.
2. Criteria of validity a. the substrate blank must be OD < 0.25
b. the negative control must be negative
c. quantitative ELISA: the mean OD-value of the standard serum must be within the validity range , which is given on the lot specific quality control certificate of the kit (after subtraction of the substrate blank!)
d. qualitative ELISA: the mean OD-value of the positive control must be within the validity range, which is given on the lot specific quality control certificate of the kit (after subtraction of the substrate blank!)
The variation of OD-values may not be higher than 20%.
If these criteria are not met, the test is not valid and must be repeated.
3. Calculation ELISAAspergillus fumigatus IgA (quantitative) - Non-automated evaluation For the test evaluation a standard curve and an evaluation table are included in the test kit so that the obtained OD-values may be assigned to the corresponding antibody activity. The reference value and the validity range of the standard serum is given on the evaluation table (quality control certificate).
The blank (A1) must be subtracted from all OD-values prior to the evaluation.
Method 1: Qualitative Evaluation To fix the cut-off ranges please multiply the mean value of the measured standard-OD with the numerical data of the certificate of quality control (see special case formulas), e.g.:
OD = 0.502 × MW(STD) with upper cut-off
OD = 0.352 × MW(STD) with lower cut-off
If the measured mean absorbance value of the standard serum is 0.64, the range of the cutoff is in between 0.225-0.321.
Method 2: Continuous determination of antibody activities using the standard curve. So called interassay variations (day to day deviations and laboratory to laboratory deviations) are compensated by multiplication of the current measured value obtained with a sample with the correction factor F. This factor is calculated as follows:

The procedure is necessary to adjust the current level of the test of the user with the lotspecific standard curve.
First, daily deviations are to be corrected by calculating a factor (correction factor F):
1. The mean of the two OD-values of the standard serum has to be calculated and checked that it is within the given validity range.
2. Calculation of the factor "F": the given reference value is divided by the mean of the extinction of the standard serum:
F = reference value extinction standard serum / mean value extinction standard serum. 3. All measured values of samples are multiplied by "F".
4. Antibody activities in IU/ml or U/ml can be determined from the standard curve with the corrected values.