Intended Use
The Trastuzumab rezetecan (ADC) ELISA Kit is a sandwich enzyme-linked immunosorbent assay (ELISA) intended for the quantitative determination of trastuzumab rezetecan (SHR-A1811), a HER2-directed antibody–drug conjugate, in human serum and plasma. The kit is designed to support pharmacokinetic (PK) and bioanalytical research on trastuzumab rezetecan.
This product is for research use only (RUO) and is not intended for use in diagnostic or therapeutic decision-making procedures.
Contents of Kit
1. Antibody pre-coated 96-well microplate (12 × 8-well strips). 1 plate, 2-8 °C
2. Trastuzumab rezetecan Standard. 3 × 20 μL,(50 μg/mL), -80 °C. Freeze–thaw is not permitted. Diluted standards are stable for 3 days at 2–8 °C.
3. HRP-conjugated Detection Antibody (100×). 1 × 120 μL, -20 °C
4. Assay Diluent, 2 × 50 mL, 2-8 °C
5. Wash Buffer (20× concentrate). 1 × 50 mL, 2-8 °C
6. TMB Substrate Solution. 1 × 12 mL, 2-8°C. Protect from light
8. Stop Solution. 1 × 7 mL, 2-8 °C
Storage
Standard store at -80°C. HRP-Conjugated Antibody Solution(100×) store at -20°C.
Other reagents store at 2-8°C.
The unopened kit is stable for 12 months.
Precision
Serum spiked 10 ng/mL: n=20, CV=5.7%
Serum spiked 1 ng/mL: n=20, CV=4.3%
Detection Range
Quantitative range: 0.78 – 50 ng/mL. 4PL fit R² ≥ 0.999.
Sensitivity
LoD: 0.34ng/mL (Based on 20 replicates of the zero standard.)
LLOQ: 0.78 ng/mL(n=20,CV=15%, Mean recovery=107%)
ULOQ: 50 ng/mL.
Note: When 20 replicates of the 0.5 ng/mL sample were assayed, the coefficient of variation (CV) of the calculated concentrations was 21%, slightly exceeding the generally accepted limit of CV ≤ 20%. Users may establish and verify acceptance criteria based on their own application requirements.
Standard Curve
The following data were obtained with the recommended procedure (single run, duplicate wells). These data are provided as an example only and must not be used in place of a standard curve generated in each individual assay run.

Hook effectNo high-dose hook effect was observed at concentrations up to 12,800 ng/mL.
Citations
Publication ()
Have you cited DEIA-JY25398 in a publication?