Contents of Kit
1. ELISA strips (colourless) coated with specific antibody [STRIPS Ab]: 1 microplate
2. Standard A [STANDARD A]: 0.6 mL, 0 ug/L, ready to use, 1 vial
3. Standard B [STANDARD B]: 0.6 mL, 0.1 ug/L, ready to use, 1 vial
4. Standard C [STANDARD C]: 0.6 mL, 0.5 ug/L, ready to use, 1 vial
5. Standard D [STANDARD D]: 0.6 mL, 1 ug/L, ready to use, 1 vial
6. Standard E [STANDARD E]: 0.6 mL, 2.5 ug/L, ready to use, 1 vial
7. MC-LR labeled with horseradish peroxidase [CONJ]: 2.5 mL, ready to use, 1 vial
8. Wash buffer concentrate, 10x concentrated [WASH 10X]: 55 mL, 1 vial
9. Dilution buffer [DIL-MC]: 15 mL, ready to use, 1 vial
10. Chromogenic substrate (TMB substrate) [TMB-BF]: 13 mL, ready to use, 1 vial
11. Stop solution [STOP]: 13 mL, ready to use, 1 vial
12. Instruction manual
Storage
Store the kit reagents at +2 to +10°C, in a dry place and protected from the light. Avoid freezing. Expiration date is indicated at the ELISA kit label and at all reagent labels.
Store unused strips in the sealable pouch and keep the desiccant inside. Transport in thermo bags until 72 hours. Any damages of packaging of kit reagents advise to the producer without delay.
Do not store diluted samples and diluted MC-LR-Px conjugate. Always prepare fresh.
Performance Characteristics
##Validity of the test##
The results of the test are valid if:
The B/Bmax of MC-LR standard ST E is not higher than 50%.
The mean absorbance of MC-LR standard ST A is not lower than 0.700.
The mean B/Bmax of Standards can be lined up as follows ST E < ST D < ST C < ST B < ST A