Class Specificity:Human serum samples previously identified as either Negative, Equivocal, or Positive in the SARS-CoV-2 IgG ELISA were incubated with either diluent alone (Untreated), diluent + goat polyclonal anti-human IgG to deplete IgG, or diluent + goat polyclonal anti-human IgM to deplete IgM, and then RBD-specific IgG binding was measured in the SARS-CoV-2 IgG assay. All samples were also evaluated in an assay for anti-SARS-CoV-2 IgM binding.

The results demonstrate that depletion of IgG (>99% reduction in IgG assay ratio) from Positive samples converts them to Negatives and that depletion of IgM (>97% reduction in IgM assay ratio, data not shown) from Positive samples has no effect on their disposition in the assay. Similarly, the Negative and Equivocal samples saw no change in assay disposition with IgM depletion, while the Equivocal sample was converted to Negative with depletion of IgG.
These data show that signal in the SARS-CoV-2 IgG ELISA is due specifically to the IgG component of a patient's immune response to SARS-CoV-2 without significant interference from SARS-CoV-2-specific IgM.
Analytical Sensitivity:Samples procured from a commercial source from 34 distinct patients that had tested positive by molecular test and tested as reactive by the Siemens ADVIA Centaur Sars-CoV-2 Total assay performed by an independent clinical laboratory were evaluated with the SARS-CoV-2 IgG ELISA.

Serum samples collected prior to December 2019 from 161 distinct donors, from two commercial sources were evaluated. One hundred fifty-eight samples (98.1%) were negative and three (1.9%) were not negative for IgG to SARS-CoV-2.
Matrix Equivalency:Serum, citrate plasma, EDTA plasma, and heparin plasma were each collected from five individuals testing negative in both SARS-CoV-2 IgG ELISA and SARS-CoV-2 IgM ELISA. Convalescent serum from an individual who tested positive in SARS-CoV-2 IgG ELISA was spiked into a 1:100 dilution of each negative matrix at four different concentrations intended to yield negative, low positive (two different concentrations), and moderate positive IgG ELISA outcomes. 1:100 negative matrices were also tested without any spike. Each combination of negative matrix and spike was tested in duplicate.
Deming regression (α = 0.05) was performed for the comparison of plasma to serum and showed the following results:

For each spike amount, there was 100 % agreement of assay outcome across all matrices, all individuals, and all replicates, thus demonstrating robust matrix equivalency.