Unit of Measure The unit of measurement for the Active-B12 (Holotranscobalamin) assay is pmol/L.
Measuring Interval (Reportable Range) The measureable range of the assay is 10 pmol/L to 128 pmol/L.
AccuracyA correlation study was performed with serum specimens from apparently healthy adults. All specimens were analysed using the Active-B12 (Holotranscobalamin) ELISA and another commercially available Holotranscobalamin assay according to the CLSI document EP9-A211. Specimen concentrations ranged from 13.8 to 112.8 pmol/L in the assay. The data obtained gave the following statistical values:
Limit of Blank In a representative study, Limit of Blank determinations were performed using two low-level holotranscobalamin samples and two reagent lots (120 replicates per reagent lot). The limit of blank of the Active-B12 (Holotranscobalamin) ELISA was found to be 4.9 pmol/L (rounded to 1 decimal place).
High Dose Hook High dose hook is a phenomenon whereby very high level specimens may read within the dynamic range of the assay. For the Active-B12 (Holotranscobalamin) ELISA. No high dose hook effect was detected with two samples with a concentration of approximately 419 and 2236 pmol/L.