Comparative Study An in-house comparative study was performed to demonstrate the equivalence of the Cardiolipin IgM ELISA (DEIA1709) test system to another commercially available Cardiolipin IgM ELISA test system. Performance was evaluated using 260
specimens as described in Table 1 below. The results of the investigation have been summarized in Table 2 below.
Table 1. Summary of Clinical Specimens

Table 2. Calculation of Relative Sensitivity, Specificity, and Agreement

Relative Sensitivity = 44/47 = 93.6%
95% Confidence Interval* = 86.6 to 100%
Relative Specificity = 163/213 = 76.5%
95% Confidence Interval* = 70.8 to 82.2%
Relative Agreement = 207/ 260 = 79.6%
95% Confidence Interval* = 74.7 to 84.5%
* 95% confidence intervals calculated using the exact method.